The biological collection efficiency of the active multi-nozzle impactor air sampler TRIO.
BAS MINI was compared to the efficiency of the slit to agar air sampler called CASELLA type
that is used as traditional reference air sampler device. The TRIO.BAS biological efficiency
was 82% vs Casella microbial air sampler [1,2]. Considering that the evaluation of biological
efficiency between different microbial air samplers is always difficult and relative, the two
results can be considered statistically comparable.
Active sampling, Agar desiccation, Airborne particles, Airborne microbes, Air flow rate,
Airborne microbial concentration, Agar medium, Bacterial cell, Biological agent, Bio sampler,
Biological efficiency, Collection efficiency, Collection surface, Culture plate, Cut-off size, d50,,
Effective collection efficiency, Experimental d50, Physical collection efficiency, Fungal spore,
Impactor, Impact on agar medium, Impaction velocity, Jet to plate distance, Microbe-carrying
particles, Microbial air sampler, Microbial Air Sampling efficiency, Environmental microbial
population, Passive sampling, Particles, Petri dish, Physical Air Sampling efficiency, Multi
nozzles impactor, Stokes number, Theoretical d50, Viable microorganism, Viable particles.
Casella type slit–to–agar sampler
Large Petri dishes (14.5 cm in diameter) are filled with agar medium to within 5 mm. of the top of the dish, (approx. 200ml) and allowed to cool at room temperature. Tryptic Soy Agar (TSA) medium is used to collect total bacteria. The air flow is 28 l/min [3]. The sampler works by drawing air through the narrow slit positioned 0.2 cm above the surface of the agar plate. While the air is being drawn through the air sampler, the plate is rotated through 360° so that the captured microorganisms are evenly distributed over the surface of TSA agar medium.
TRIO.BAS Sieve Impaction to agar active Sampler
A known volume of air is aspirated inside the chamber of the sampler. The microorganisms are captured by impact on the agar surface of a plate (Contact Plate or Petri dish). Different media are used for different micro-organisms (i.e.: Sabouraud Dextrose Agar (SDA) for yeast and molds, Tryptic Soy Agar (TSA) for total bacterial count). The air flow is set to 100 l/min (Figure 1-2).
Figure 1: “Slit to agar” air sampler and “Sieve Impaction Active agar” air samplers
The comparative tests were performed in an indoor room (7.0x 4.5x 4.0 m.) with doors and windows closed with the purpose to evaluate the microbial population in the room’s atmosphere [4-6]. No people were present during the tests.
The two air samplers were positioned in the center of the room at a distance of 1 m and 70 cm high. Both instruments were used according to the instructions of the manufacturers.
The aspirated volume of air was 1.000 l. (38 minutes for Casella sampler; 10 minutes– Interval fractioned sampling 3x 3x 3 - for TRIO.BAS). The comparison was therefore considering about 1.000 l. of aspirated air.
For both devices the culture media was the TSA (Trypticase Soy Agar with Lecithin and Polysorbate 80, in Petri dish manufactured by BD). The incubation time was for 48 hours at the temperature of 32°C. The numbers of Colony Forming Units (CFU) were counted manually
Triplicate tests were performed for seven consecutive days and the results are reported
in table 1 as CFU /Petri dish / 1.000 l. of air
The biological efficiency was monitored as the comparative efficiency of collection of normal indoor microbial population in a closed room of house.
The biological efficiency of the air sampler “A” TRIO BAS was 82%
vs air sampler “B” Casella. Considering that the biological efficiency
between different air samplers is always difficult and relative, the two
results can be considered statistically comparable.